Description
AAS –series
Specifications:
Wavelength range | 190-900 nm |
Wavelength accuracy | ±0.25nm |
Resolution | Two spectral lines of Mn at 279.5nm and 279.8nm can be separated |
with the spectral bandwidth of 0.2nm and valley-peak energy ratio less than 30%. | |
Baseline stability | ≤0.004A/30min |
Background correction | The D2 lamp background correction capability at 1A is better than 30 times. |
The S-H background correction capability at 1.8A is better than 30 times. (only for AAS-10B/12B) |
Hollow Cathode Lamps:
Lamp turret | 6-lamp turret (AAS-10B/12B), 4-lamp turret (AAS-13B) |
Auto-alignment, fully automated scan and peak-picking. | |
Lamp current adjustment | Automatic adjustment and displa. |
Wide pulse current: 0~25mA | |
Narrow pulse current: 0~10mA | |
Lamp power supply mode | 400Hz square wave pulse |
100Hz Narrow square wave pulse + 400Hz wide square wave pulse (AAS-10B/12B) |
Optical System:
Monochomator | Single beam, Czerny-Turner design grating monochromator |
Grating | 1800 l/mm |
Focal length | 277mm |
Blazed Wavelength | 250nm |
Spectral Bandwidth | 0.1nm, 0.2nm, 0.4nm, 1.2nm automatic change. |
Flame Atomizer:
Burner | 10cm single slot all-titanium burner |
Spray chamber | Corrosion resistant all-plastic spray chamber. |
Nebulizer | High efficiency glass nebulizer with metal sleeve, sucking up rate: 6-7mL/min |
Emission burner | Provided with AAS-10B/12B |
Detection and Data Processing System:
Detector | R928 Photomultiplier with high sensitivity and wide spectral range. |
Software | Windows operating system |
Analytical method | Working curve auto-fitting; standard addition method; automatic sensitivity correction; |
automatic calculation of concentration and content. | |
Repeat times | Maximum 20 times of repeat measurement, automatic calculation of mean value, |
standard deviation and relative standard deviation. | |
Multi-task function | Sequential measurement of multi-elements in one sample |
Condition reading | With model function |
Result printing | Measurement data and final analytical report printout, editing with Excel. |
Port | Standard RS-232 serial port communication |
Characteristic Concentration and Detection Limit:
Normal Air-C2H2 flame | Cu: Characteristic concentration≤0.025mg/L, Detection limit≤0.006mg/L; |
Oxygen-rich | Ba: Characteristic concentration ≤ 0.22mg/L |
Air- C2H2 flame | Al: Characteristic concentration ≤ 0.4mg/L (for AAS-10B) |
Function Expansion:
Hydride vapor generator can be connected for hydride analysis
Dimensions and weight:
Main unit (LxWxH) | 1020 x 490 x 540 mm / Net weight : 80kg |
Ordering Information:
Model | Description of Article |
AAS-13B | 4-lamp turret, Air-C2H2 flame, Hollow Cathode Lamps : Cu, Mn, Hg |
The D2 lamp background | |
AAS-12B | 6-lamp turret, Air-C2H2 flame, Emission burner, Hollow Cathode Lamps : Cu, Mn, Hg |
D2 & S-H background correction | |
AAS-10B | 6-lamp turret, Air-C2H2 flame and patented air-C2H2-O2 flame, Emission burner |
D2 & S-H background correction, Hollow Cathode Lamps : Cu, Mn, Hg, Ba | |
Al: Characteristic concentration ≤ 0.4mg/L (for AAS-10B) |
Accessories on Request |